9.13 Setting the 8940 F/V UNIT
────────────────────────────────────────────────────
Frequency Obtains the frequency of the input pulse corresponding to
the measured waveform.
Count Counts the number of input pulses.
Pulse Duty Ratio Obtains the duty ratio of the measured waveform.
Voltage Measures the waveform voltage.
Current Measures the waveform current.
NOTE
5V
-5 V
0V
Threshold
Threshold Setting Example
9
.13 Setting the 8940 F/V UNIT
The 8940 F/V UNIT can measure frequency, count, pulse duty ratio, voltage
and current on each channel (with one measurement type per channel).Refer
to Section 9.3 for the common settings.
■Measurement Mode Selection
Select the measurement mode from the following:
・Hold and pull up settings cannot be determined from the DISPLAY or
CHANNEL screen. Check through the function display that appears when
the flashing cursor is moved to the "range" item.
・During list printing, Hold and Pull-Up on/off selections are printed. Refer to
Section 13.5 for details.
・The input coupling is fixed at DC for Frequency, Count and Pulse Duty
Ratio measurement modes.
・Measurement results may differ depending on the threshold setting. To
obtain the correct measurement, set the threshold to match the input
waveform.
・Probe check function is available only when using the 9322
DEFFERENTIAL PROBE. For details on operating procedure, refer to the
"9322 DEFFERENTIAL PROBE" Instruction Manual.