9.12 Setting the 8939 STRAIN UNIT
────────────────────────────────────────────────────
NOTE
・Connect the sensor, make sure there is no input, and execute auto
balancing.
・Auto balancing is disabled during measurement.
・The key manipulation is rejected during measurement.
・Repeat the auto balancing when the input unit was changed, when the
power is on/off, or when the system is set to reset, when the measurement
range is change, when the strain gauge adapter is change.
・When there is a sudden change in ambient temperature the zero position
may drift. To assure continued measurement precision, perform the auto
balancing again.