User Manual
238
────────────────────────────────────────────────────
9.12 Setting the 8939 STRAIN UNIT
────────────────────────────────────────────────────
Screen: CHANNEL1, DISPLAY
Call up the CHANNEL1 or
DISPLAYscreen.
Use the CH.SET key to display the desired
channel screen.
Move the flashing cursor to the position
shown in the figure on the left.
Use the function keys, the JOG control or
the RANGE key to make the selection.
Method
Method
(1) Setting the measurement range
:Move the arrow up in the selection
window.
:Move the arrow down in the
selection window.
Selection
Selection
(2) Setting the auto balancing
Move the flashing cursor to the point
shown in the figure on the left.
Use the function keys to set the auto
balancing.
:All strain channels execute the auto
balancing.
:Only selected channel execute the
auto balancing.
Selection
Selection
Flashing curso
r
9
.12 Setting the 8939 STRAIN UNIT
For the channel to which the 8939 STRAIN UNIT is assigned, an auto
balancing must be executed through the range setup menu.
Auto balancing is a function by which the reference output level of the
converter is adjusted to the specified zero position.
Before the auto balancing is performed, the strain unit must be warmed up for
approximately 1 hour after power-on to stabilize its internal temperature.
The scaling function can be used to convert an output voltage from a sensor or
similar into a physical quantity. See Section 9.7.2.
Refer to Section 9.3 for the common settings.