User Manual

235
────────────────────────────────────────────────────
9.10 Setting the 8937 VOLTAGE/TEMP UNIT
────────────────────────────────────────────────────
Move the flashing cursor to the position
shown in the figure on the left.
Use the function keys to set the drift
compensation.
Selection
Selection
Flashing curso
r
(3) Setting the drift compensation
Drift compensation is a function that periodically
(about once per second) cancels the variances
in the reference voltage that accumulate over
time.
Drift compensation is enabled.
Drift compensation is disabled.
Drift compensationON
Move the flashing cursor to the position
shown in the figure on the left.
Use the function keys to set the reference
junction compensation.
Selection
Selection
Flashing curso
r
(4) Setting the reference junction compensation
Reference junction compensation is
performed internally by the 8937.
(Measurement accuracy is the sum
of the accuracy of temperature
measurement plus reference
junction compensation accuracy.)
Reference junction compensation is
not performed by the 8937. Select
this option when using an external
reference junction compensation
unit.
(Measurement accuracy refers only
to the accuracy of temperature
measurement.)
Internal: No underbar appears below C
.
External: The underbar appears below C
.
Discriminating between internal and
external reference junction compensatio
n