9.10 Setting the 8937 VOLTAGE/TEMP UNIT
────────────────────────────────────────────────────
WARNIN
G
A common GND is used for voltage and temperature input on all channels
.
Never input voltage and temperature simultaneously, since doing so could
result in damage to the sample being tested.
NOTE
Voltage Measures the waveform voltage.
Temperature
TMP
-
Measurement range TMP
-
Measurement range
K -200 to 1350 N -200 to 1300
E -200 to 800 R 0 to 1700
J -200 to 1100 S 0 to 1700
T -200 to 400 B 300 to 1800
9
.10 Setting the 8937 VOLTAGE/TEMP UNIT
Digital filter and dfift compensation settings cannot be determined from the
DISPLAY or CHANNEL screen. Check through the function display that
appears when the flashing cursor is moved to the "range" item.
Measurement conditions for the measured waveform data can be determined
from the printed listing. The settings for the digital filter, drift
compensation, reference contact compensation and the type of thermocouple
can be determined. See Section 13.5.
The 8937 VOLTAGE/TEMP UNIT allows measurement of voltage or
temperature on any channel.
Refer to Section 9.3 for the common settings.