9.9 Setting the Display Screen
────────────────────────────────────────────────────
9
.9.2 Vernier Function
Press the DISP key to call up the
DISPLAY screen.
Use the CH.SET key to display the desired
channel screen and use the cursor key to
move the flashing cursor to the desired
channel.
Use the function keys to select the vernier
function.
Use
to magnify and to compress.
The adjustment range is from 1/2 to 2
times of the original waveform.
Method
Method
Displays the magnified waveform.
×
Displays the waveform with no
vernier scaling.
Displays the compressed waveform.
Cancels the vernier function. (The
original waveform returns.)
Magnifies the waveform.
Compresses the waveform.
Selection
Selection
Flashing curso
r
NOTE
Using fine adjustment, the input voltage can be matched to a desired
reading.
For example, an actual input voltage of 1.8 V can be converted to a 2.0 V
reading.
The vernier function is not applicable to a waveform after waveform
processing.
The ratio of the waveform’s enlargement or compression is not displayed.
The vernier function ON/OFF setting can be checked after the CH.SET key
is pressed to display each channel setting on the screen. This setting cannot
be checked using the printout or list print function.