User Manual

216
────────────────────────────────────────────────────
9.7 Scaling Function (SYSTEM2)
────────────────────────────────────────────────────
9
.7.1 Setting the Scaling Function
Screen: SCALING (SYSTEM2
)
Press the SYSTEM key to call up the
SYSTEM2 screen.
Move the flashing cursor to the position
shown in the figure on the left.
Use the function keys to make the
selection.
Method
Method
Use conversion ratio method
Use 2-point method
Selection
Selection
Use the CH.SET key to display the desired
channel screen.
Use the cursor key to move the flashing
cursor to the desired channel.
Use the function keys to make the
selection.
Scaling not used
Scaling used
(specify exponent as integer)
Selection
Selection
Scaling used
(specify exponent as multiple of 3)
Reset.
(1) Select the Scaling method
(2) Setting the scaling function
Flashing curso
r
Flashing curso
r
NOTE
If the exponents of data are integers (SCI), the numeral used as the cursor
value or the value to be printed out will be expressed as it is rather than in