User Manual
209
────────────────────────────────────────────────────
9.5 Zero Setting
────────────────────────────────────────────────────
9
.5.2 Zero Offset
Screen: CHANNEL1, DISPLAY
Call up the CHANNEL1 or DISPLAY
screen.
Move the flashing cursor to the
zero pos. item.
Press the F5 [ 0 Setting ].
Method
Method
Flashing curso
r
Use the function keys to make the
selection.
:Displays the Zero Offset and Zero
Adjust setup screen.
Selection
Selection
Selection
Selection
:Activates zero offset on all
channels.
:Activates zero offset on specified
channels.
:Disables zero offset on specified
channels.
:Exits Zero setup.
:Performs zero adjustment.
Disables zero offset on all channels
.
This function offsets the input voltage measurement to 0 V to compensate for
an actual non-zero input voltage. If the actual input voltage exceeds ±10
DIVs, zero offset cannot be performed. This function is for the voltage and
current measurement only.
The zero offset is active when power is off, but if the input unit configuration