User Manual

207
────────────────────────────────────────────────────
9.4 Copying Channel Settings
────────────────────────────────────────────────────
Screen: CHANNEL1, DISPLAY
Press the CHAN key to call up the
CHANNEL1 screen.
Move the flashing cursor to the number of
the channel to be used as copy target.
Use the function keys or the JOG control
to specify the copy source channel.
Method
Method
Increase channel number
Decrease channel number
Selection
Selection
Selection of "for ALL" enables the settings
(comments) for the channel with the
flashing cursor to be copied to all channels
.
When the F5 [ set ] is pressed, the
settings of the copy source channel are
copied to the copy target channel.
For variable function and scaling function,
the settings are copied in an identical.
copy targe
t
copy sourc
e
NOTE
9
.4 Copying Channel Settings
Copies any of input channel settings (voltage axis range, input coupling, and
low-pass filter), variable function settings, scaling function settings, or
comment function settings to another channel.
Copying is not possible if the copy source and destination are different, or if
the measurement mode is different when copying from and to the same unit.
Magnification and zero position are copied to the desired destination only
when the copy source is set to "for ALL" when copying is performed in the
"input channel settings."
It is not possible to copy the settings of input channels and variable function
in a single operation.