User Manual

204
────────────────────────────────────────────────────
9.3 Setting the CHANNEL1 Screen
────────────────────────────────────────────────────
9
.3.7 Setting the Low-Pass Filter
Screen: CHANNEL1, DISPLAY
Call up the CHANNEL1 or DISPLAY
screen.
Use the CH.SET key to display the desired
channel screen.
Move the flashing cursor to the point
shown in the figure on the left.
Use the function keys to make the
selection.
Method
Method
No low-pass filter is connected.
Connect a filter with a cutoff
frequency of 5 Hz
Selection
Selection
Connect a filter with a cutoff
frequency of 500 Hz
Connect a filter with a cutoff
frequency of 5 kHz
(Ex: 8936 ANALOG UNIT
)
Connect a filter with a cutoff
frequency of 100 kHz
Flashing curso
r
Input Unit Low-Pass Filter
8936 OFF, 5, 500, 5 k, 100 kHz
8937 Voltage OFF, 5, 500, 5 k, 100 kHz
Temperature OFF, 5, 500Hz
8938 OFF, 5, 500, 5 k, 100 kHz
8939 OFF, 10, 30, 300, 3 kHz
8940 OFF, 5, 500, 5 k, 100 kHz
8947 Voltage OFF, 5, 500, 5 k, 100 kHz
Charge, Pre-AMP OFF, 500, 5 kHz
Low-pass filters internal to the input units are set.
Effective for removing unneeded high-frequency components.
Types of Low-Pass Filters in Each Unit