User Manual

195
────────────────────────────────────────────────────
9.2 Input Channel Setting Procedure
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
入力チャネルの設定を行います
SYSTEM Screen
(Advanced function)
Two types of comment are available.
(1) Title comment input
(2) Input of comments for each channel
For details on comment function, refer to Section 9.8.
入力チャネルの設定を行います
Zero settings
After the power switch is turned on, wait approximately 30 min for the
8936, 8938, 8940 and approximately 1 h for the 8937, 8939, 8947 to
stabilize the inside temperature of the connected input unit in order to
obtain accurate waveforms. Then, make a zero adjustment of the 8936,
8937, 8938, 8940 or 8947 perform an auto balancing of the 8939 prior to
measurement.
Zero adjustment See Section 9.5.1.
Auto balancing See Section 9.12.
入力チャネルの設定を行います
DISPLAY Screen
Direct channel setting can be performed using the CH&NUM
key, the POSITION key, and the RANGE key.
Pressing the CH.SET key, input channel setting can be set on
DISPLAY screen. See Section 9.9.
入力チャネルの設定を行います
CHANNEL2 Screen
The variable function allows the user to modify the waveform position
and size. For details on variable function, refer to Section 9.6.1.
(Advanced function)
The scaling function can be used to convert an output voltage from a
sensor or similar into a physical quantity. For details on scaling
function, refer to Section 9.7.
Variable Function
SYSTEM2SCALING
SYSTEM3COMMENT
Zero ajustment
Zero Offset
This function offsets the input voltage measurement to 0 V to
compensate for an actual non-zero input voltage.
See Section 9.5
.
NOTE
The "Input Channel Setting Procedure" corresponds to the 8936 ANALOG
UNIT description.
Refer to each item for particular descriptions related to each input unit.
8937 VOLTAGE / TEMP UNIT
Refer to Section 9.10.
8938 FFT ANALOG UNIT
Refer to Section 9.11.
8939 STRAIN UNIT
Refer to Section 9.12.
8940 F/V UNIT
Refer to Section 9.13.
8947 CHARGE UNIT
Refer to Section 9.14.