8.6 Start and Stop Operation (FFT)
────────────────────────────────────────────────────
・See Section 8.3 STATUS1 settings
・See Section 8.4 STATUS2 settings
・See Chapter 9 Input channel settings
・See Chapter 10 Trigger Functions
・Data recording starts when trigger condition are met.
・When trigger mode selected AUTO, unit waits for about 1 second
for trigger conditions to be met. After this interval, data recording
starts, regardless of trigger state.
・Press the START key and LED light.
・When the pre-trigger is set, the trigger will not be registered for a
certain period after the start of measurement. (During this interval,
Pre-trigger standby is shown on the display.)
・When the trigger can be registered, the indication Waiting for trigger
is shown on the display.
・Displays "Storing".
・When roll mode is set to ON, the waveform is displayed (scrolling
)
immediately at the start of recording
・
・When the STOP key is pressed twice during measurement, the
8826 is forcibly stopped.(Auto printout and auto save are not
executed.)
・Waveform displays after data corresponding to samping point have
been stored in memory.
・
Auto printout and auto save are executed.
・End of measurement
・Each time when trigger conditions are met, data are recorded and
memory contents are overwritten.
【 End of measurement in REPEAT and AUTO trigger modes 】
・When the STOP key is pressed once during measurement, the
8826 acquires measurement data in an amount corresponding to
the set recording length, and the measurement is stopped.
(Waveform display, auto printout, and auto save are executed.)
Trigger condition are
met
Setting to the 8826
・STATUS Settings
・Channel Settings
・Trigger Settings
Setting the 8826
・STATUS Settings
・Channel Settings
・Trigger Settings
Measurement Start
Measurement Start
Recording Start
Measurement end
Measurement end
Recording end
Recording end
Trigger mode:SINGLE
Trigger mode:REPEAT
AUTO
8
.6 Start and Stop Operation (FFT)