8.4 STATUS2 Settings (FFT)
────────────────────────────────────────────────────
8
.4.6 Setting the Waveform Evaluation
Screen: STATUS
2
Press the STATUS key to call up the
STATUS2 screen.
Move the flashing cursor to the
wave comparison item, as shown in the
figure on the left.
Use the function keys to make the
selection.
Disable waveform evaluation.
Disable waveform evaluation.
Return NG if any part of the
waveform leaves the evaluation
area.
Selection
Selection
Method
Method
Return NG if the entire waveform
leaves the evaluation area.
Evaluation area is created.
NOTE
The Waveform Evaluation function can be used single screen and Nyquist
screen.
GO (pass) or NG (fail) evaluation of the input signal waveform can be
performed using an evaluation area specified by the user.
This can serve to detect irregular waveforms.
After the evaluation result is generated, signals are output from the GO/NG
terminal.
Only one waveform evaluation area is saved in the internal memory. For
example, when the memory recorder function is changed to the FFT function
and the waveform evaluation area of the FFT is saved, the waveform
evaluation area created using the memory recorder function is erased.
For details on the waveform evaluation, refer to Section 16.3.