User Manual
157
────────────────────────────────────────────────────
8.3 STATUS1 Settings (FFT)
────────────────────────────────────────────────────
8
.3.11 Setting the Display Scale
Screen: STATUS
1
Press the STATUS key to call up the
STATUS1 screen.
Move the flashing cursor to the scale item,
shown in the figure on the left.
Use the function keys to make the selection
.
Method
Method
:The vertical axis (Y-axis) scale is
set automatically, depending on the
processing result.
:The vertical axis (Y-axis) scale can
be set as desired, to match the
purpose of the measurement.
This is useful for enlarging or
reducing the amplitude and for
shifting the waveform up or down.
Selection
Selection
When MANUAL is selected
・The upper and lower limits for the display
scale can be set by the user.
・Make settings to match processing results.
・Setting range is -9.9999E+29 to
9.9999E+29. (exponent is E-29 to E+29).
Move the flashing cursor to the lower item.
Use the JOG control, the function keys or
CH&NUM key to select the lower limit for
each digits.
Set the upper limit in the same manner.
:Increases the value.
:Decreases the value.
Selection
Selection
:Use CH&NUM key to enter the
value.
The display scale for showing the FFT processing result can either be set
manually or automatically.