User Manual
136
────────────────────────────────────────────────────
7.3 STATUS1 Settings (REC&MEM)
────────────────────────────────────────────────────
Move the blinking cursor to "Memory
Recording Length", and select using the
jog control or function keys.
:Move the cursor up in the selection
window.
:Move the cursor down in the
selection window.
Selection
Selection
NOTE
・The sampling cycle of the recorder is the same as that set for the memory,
so depending on the sampling cycle, there are some time axes that cannot be
set for the recorder. Refer to Section 7.3.2 for details.
・The recordable time (recording length) is determined by the space available
on the PC card or MO disk and the number of channels being saved.
・The recording length of the recorder is determined automatically by the
recorder time axis range, memory time axis range and memory recording
length.
・Channels cannot be selected for real-time save (all displayed channel data
waveforms are saved).
・Partial saving (only between cursors) is not possible even when the A-B
cursors are displayed.
・When Real-Time Save is enabled, Auto Save, Auto Print, Additional
Recording and Sequential Save settings are ignored.
・When measurement is aborted before the set recording length, the
waveforms input prior to the abort are saved.
・Files with extensions MEM, REC and R_M are created during Real-Time
Save. Refer to Section 14.7, "SAVE" for file name details.
・Real-Time Save cannot save data in text format ("Text Save").