User Manual
132
────────────────────────────────────────────────────
7.3 STATUS1 Settings (REC&MEM)
────────────────────────────────────────────────────
7
.3.10 Setting the Auto Save Function
Screen: STATUS
1
Press the STATUS key to call up the
STATUS1 screen.
Move the flashing cursor to the auto save
item, as shown in the figure on the left.
Use the function keys to make the
selection.
:Auto save is disabled.
:Automatically stored on floppy disk
Selection
Selection
Method
Method
:Automatically stored on PC card.
:Automatically stored on connected
SCSI device.
:Automatically stored on MO disk.
When Media is selected, the Storage
method item appears.
:When the media becomes full,
automatic storage stops.
:When the media becomes full, old
files are deleted to make room for
automatic storage.
Selection
Selection
When 1 BLOCK save is selected with the Binary Format, the
memory waveforms with the file extension ’MEM’ and the
recorded waveforms in the ’REC’ file is deleted.
When ALL BLOCKS save is selected with the Binary Format,
files with the extensions ’R-M’, ’MEM’ and ’REC’ are deleted.
With the text format selected, files with the extension ’TXT’ are
deleted.
:Data can only be transferred to the
9333 LAN COMMUNICATOR when
using a LAN card.
・When the function is enabled, measurement data are automatically stored
on a floppy disk, PC card, MO disk or connected SCSI device after they are
captured.
・The Auto Save function stores a file in the directory currently selected on
the file screen. See Section 14.12.2.