User Manual

115
────────────────────────────────────────────────────
6.5 Start and Stop Operation (RMS)
────────────────────────────────────────────────────
See Section 6.3STATUS settings
See Chapter 9 Input channel settings
See Chapter 10Trigger Functions
Data recording starts when trigger condition are met.
Press the START key and LED light.
When the trigger can be registered, the indication Waiting for trigger
is shown on the display.
Displays "Storing".
Waveform appears on the screen and printout starts.
Using the VIEW key to display the time (from measurement starte
d
time).
When Auto Save is on, pressing the STOP key during
measurement causes measurement data to be saved automaticall
y
until the stop.
End of measurement
Each time when trigger conditions are met, data are recorded and
memory contents are overwritten.
End of measurement in REPEAT trigger modes
When STOP key is pressed, measurement stops after data
corresponding to recording length have been stored in memory.
(auto save are not executed.)
Trigger condition are
met
Setting to the 8826
STATUS Settings
Channel Settings
Trigger Settings
Setting the 8826
STATUS Settings
Channel Settings
Trigger Settings
Measurement Start
Measurement Start
Recording Start
Measurement end
Measurement end
Recording end
Recording end
Trigger mode:SINGLE
Trigger mode:REPEAT
Auto save are executed.
6
.5 Start and Stop Operation (RMS)