6.3 STATUS Settings (RMS)
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
6
.3.2 Setting the Time Axis Range
Screen: STATUS, DISPLAY
Call up the STATUS or DISPLAY screen.
Move the flashing cursor to the time/div
item, as shown in the figure on the left.
Use the JOG control, the function keys or
TIME/DIV key to make the selection.
Method
Method-1
Method
Method-2
Call up the STATUS or DISPLAY screen.
Use the TIME/DIV key to make the
selection.
The TIME/DIV key can be used regardless
of where the flashing cursor is located.
:Use the TIME/DIV ke
y
NOTE
6
.3.3 Setting the Frequency
Screen: STATUS
Press the STATUS key to call up the
STATUS screen.
Move the flashing cursor to the frequency
item, shown in the figure on the left.
Use the function keys to make the
selection.
Method
Method
:Measure rms value of 50 Hz signal
:Measure rms value of 60 Hz signal
Selection
Selection
・Set the speed for inputting and storing the waveform of the input signal.
・Time axis range setting expresses the time for 1 DIV.
・The sampling interval is constant (200μs), regardless of the time axis range.
For details on sampling period, refer to Section 21.3.5.
Serves to set the frequency of the signal to be measured.