User Manual
96
────────────────────────────────────────────────────
6.1 Outline of the RMS Recorder Function
────────────────────────────────────────────────────
6
.1 Outline of the RMS Recorder Function
The RMS recorder function has the following features.
(1) The voltage value and DC signal for the commercial power supplies are displayed
and recorded as the rms value.
(2) Real time continuous recording
(3) The 8826 is designed to measure commercial power supplies (50/60 Hz) and DC
signal.
(4) RMS accuracy:±3%f.s.
(5) Time axis setting
5 s/DIV to 1 h/DIV
(6) Time axis resolution: 100 points/DIV (printer)
(7) Sampling speed
20 rms data/s (200 μs fixed)
(8) Memory capacity
・16 M words: 2000 DIV
・64 M words: 10000 DIV
・Arbitrarily (Set from 1 division to the maximum number of divisions at 1-
division intervals)
(9) Waveform magnification/compression display and printout
・Time axis direction: ×1to×1/500
・Voltage axis direction: ×10 to ×1/2 (single, X-Y single)
×5to×1/4 (dual to hex, X-Y quad)
・With the variable function, vernier function
(10) Display format
Time axis waveform: single, dual, quad screen display (LCD), oct, hex screen
display(printer only)
(11) Scrollable display
・The most recent 2000 (64M words: 10000) divisions of the data are stored in
memory.
・It is possible to scroll back for easy review.
(12) Additional recording function
・The first set of measurement data is preserved, and recording of the second
set of measurement data starts after the first set.
(13) Logging function
Numeric printout of waveform data
(14) Reprint function
The most recent 2000 (64 M words: 10000) divisions of the data stored in
memory can be printed as many times as required.