5.4 Using the X-Y CONT Recorder
────────────────────────────────────────────────────
NOTE
Four single plot is displayed and recorded
.
A single plot is displayed and recorded.
(1) 1 画面
(1) X-Y Single
(1) 1 画面
(2) X-Y Quad
5
.4 Using the X-Y CONT Recorder
(1) The same operation as a normal recorder is available to plot between
channels (real time X-Y recording).
(2) Unlike an X-Y plot produced in the memory recorder function mode, the
time axis information for each channel is not being recorded.
(3) Any of channels 1 to 32 can be selected for each of the X and Y axis. Up to
four X-Y plots can be made simultaneously.
(4) For dot display the sampling period is fixed at 500 μs, while for line
display the fastest sampling period is 500μs (unfixed).
(5) There is no limit on the length of a recording because basically the
operation is the same as that of a conventional recorder.
(6) When the waveform clear is OFF, overlay can be performed.
(7) Measurement data and setting data can be saved to media.
・The waveforms with a display format that has been set to screens 1 to 16
cannot be subject to X-Y plotting following measurement. In addition, the
waveforms measured by setting their display formats to the X-Y screen
cannot be displayed on screens 1 to 16 following measurement.