User Manual
78
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
────────────────────────────────────────────────────
Screen: STATUS, DISPLAY
Call up the STATUS or DISPLAY screen.
Move the flashing cursor to the shot item,
as shown in the figure on the left.
Select the ARBITRARY.
Use the JOG control or the function keys
to make the selection. Use the cursor keys
to change the column.
:Value up.
:Value down.
:Set the fixed recording length mode
.
Selection
Selection
Method
Method-2
Any recording length mod
e
:Set the any recording length mode.
NOTE
・When the recording length is set to "CONT" within the range of 20 to 200
ms/DIV, the printer setting (real-time printing) is automatically disabled.
・When the recording length is set to "CONT," auto saving is automatically
disabled.
・The maximum recording length is 2000 divisions for a memory capacity of
16 M words or 10,000 divisions for 64 M words (with additional memory).