5.3 STATUS Settings (REC)
────────────────────────────────────────────────────
5
.3.3 Setting the Sampling Period
Screen: STATUS
Press the STATUS key to call up the
STATUS screen.
Move the flashing cursor to the sampling
item, shown in the figure on the left.
Use the JOG control or the function keys
to make the selection.
Method
Method
Move the cursor up in the selection
window.
Move the cursor down in the
selection window.
Selection
Selection
NOTE
The available range depends on the selected time axis range (input signal
waveform capture rate).
When a short sampling period is set and the input waveform changes slightly,
a sudden disturbance such as noise will increase the difference the between
the maximum and minimum values. To eliminate this phenomenon, set a
long sampling period. For details, see Section 21.3.4.