User Manual

3
______________________________________________________________________
(1) Measuring the Voltage Waveform of a 220 V AC Source (Memory Recorder Function)
________________________________________________________________
1. Function:
MEMORY
2. Time axis range:
5ms
3. Shot: optional setting
1. Press the SET key to display the Status screen.
1. Press the SET key to display the Status screen.
1
2
3
SET
DISP
SYSTEM
1. Press the SET key to display the Status screen.
2. Press the SET key to display the Channel screen.
Settings for Channel 1 (CH1)
1. Waveform color
:
optional setting
2. Voltage axis range:
100 V
(The waveform peak value
of 220 Vrms is
approximately 311 V.)
3. Zero position:
50 %
(Set the center of the
screen to 0 V)
SET
DISP
SYSTEM
1
2
3
(1) Measuring the Voltage Waveform of a
220 V AC Source
(
Memor
y
Recorder Function
)
This example describes how to measure the voltage waveform of a
220 V AC (50 Hz) commercial power source using the Memory
Recorder function.
Move the flashing cursor using the cursor keys (
), and use
the
and keys to make a setting.