(3) Measuring a Current Waveform Using the 9018-10 (Memory Recorder Function)
________________________________________________________________
1. Function:
MEMORY
2. Time axis range:
5ms
3. Shot: optional setting
1. Press the SET key to display the Status screen.
1. Press the SET key to display the Status screen.
1
2
3
SET
DISP
SYSTEM
1. Press the SET key to display the Status screen.
2. Press the SET key to display the Channel screen.
Settings for Channel 2 (CH2)
1. Waveform color
: optional
setting
2. Input coupling:
CL AMP
(unit: A)
3. Voltage axis range: Set the
range to correspond with the
current range of the 9018-10
Example: 20 A
4. Current value for one division
(along the measurement axis
range)
5. Zero position:
50%
(set the center of the screen to
0 A.)
SET
DISP
SYSTEM
1
5
2
3
4
(3) Measuring a Current Waveform Using the
9018-10
(
Memor
y
Recorder Function
)
The 8807-01, 8808-01, 8807-51, 8808-51
MEMORY HiCORDER
can measure only voltage by itself, but when connected with the
9018-10 CLAMP ON PROBE, it can also measure current.
When changing the current range of the 9018-10 CLAMP ON
PROBE, be sure to also change the voltage axis range of the 8807-
01, 8808-01, 8807-51, 8808-51 to correspond to the new range. If
the ranges of the two units do not correspond, incorrect values are
displayed.