Data Sheet
6
Basic specifications (MEMORY HiCORDER 8807-01, 8808-01)
Measurement
functions
(1) Memory recorder, (2) Recorder,
(3) RMS recorder
& memory (50/60 Hz/ or DC only)
Input type and
number of channels
8807-01: fixed input section, 2 analog + 8 logic,
8808
-01: fixed input section 4 analog + 8 logic
Isolated analog channels, isolated input and outputs, logic has common
GND.
Maximum sampling
rate
400 k sample/s (2.5 µs cycle)
Simultaneous sampling for 2/4 analog + 8 logic channels
Memory capacity
8807
-01: (analog 12 bits + logic 4 bits) × 256 kilo-words/
channel (CH1) to (analog 12 bits + logic 4 bits) × 128 kilo-
words/channel (CH1, CH2)
8808
-01: (analog 12 bits + logic 4 bits) × 256 kilo-words/
channel (CH1) to (analog 12 bits + logic 4 bits) × 64 kilo-words/
channel (CH1 - CH4)
External memory
PC card TYPE II slot × 1: flash ATA card (max. 1GB), MS-DOS
format
Memory contents: Setting conditions, measurement data (binary,
text), image data (BMP), calculation results (figures)
Battery backup
Clock, waveform data, settings, battery life approx. 5 years (at 25 ˚C/ 77 ˚F)
External control
Terminal block: trigger input/output
Interface
RS-232C interface: 9-pin round connector terminal
(the optional RS-232C CABLE 9612 is required for connection to PC)
Printer interface: PRINTER UNIT 8992 can be connected (option)
Environment
(no condensation)
Operation: +5 ˚C/41 ˚F to +40 ˚C/104 ˚F, 35% to 80% relative humidity.
Storage: -10 ˚C/14 ˚F to +50 ˚C/122 ˚F, 35% to 80% relative humidity.
Applicable standards
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
Power supplies
*
1
Note: These LR6/AA
alkaline batteries cannot be
used with the PRINTER UNIT
8992.
(1) AC Adapter model 9418-15 or 9418-10 (DC 12V ±10%)
(2) *
1
LR6/AA alkaline batteries × 6 (AC adapter has priority when
used in combination with battery pack)
(3) BATTERY PACK 9447 (AC adapter has priority when used in
combination with battery pack, fast recharge possible with AC adapter)
(4) 12 V Car battery (Please contact HIOKI for connection cord).
Power consumption
8807-01, 8808-01: 15 VA max. (when using optional printer)
Continuous
operation time
Approx. 3 hours
(when using BATTERY PACK 9447)
Approx. 1 hours (when using *
1
alkaline batteries)
Charge time
With power switch OFF, approx. 2 hours fast charge (at 23 ˚C/ 73 ˚F)
Dimensions and
mass
8807-01, 8808-01: approx. 203 (7.99) W × 170 (6.69) H × 52 (2.05) D mm (inch)
(printer detached)
8807-01, 8808-01: approx. 280 (11.02) W × 170 (6.69) H × 52 (2.05) D mm (inch)
(printer attached)
8807
-01: approx. 1.1 kg/ 38.80 oz (printer detached) 1.5 kg/ 52.91 oz (printer attached)
8808
-01: approx. 1.2 kg/ 42.33 oz (printer detached) 1.6 kg/ 56.44 oz (printer attached)
Supplied accessories
LR6/AA alkaline batteries
×6
, alkaline battery box
×1
, strap
×1
,
Application disk
×1
-Specifications-
Memory function
Time axis
200 µs to 5 minutes/division, 19 settings, time axis zoom
×2 to ×10; 3 settings, compression 1/2 to 1/500; 8 settings
Sampling period
1/80 of time axis ranges
(minimum sampling period 2.5 µs)
Recording length
20 to 3200*
3
divisions
*
3
Depending on the number of channels in use.
Pre-trigger
Can record data from before the trigger point, 0 to 100 %
or -95 % of recording length; 15 settings
Other functions
Numerical calculations, logging (numerical printout), X-Y waveform
plot (one plot on the 8807- 01, up to three plots on the 8808-01), voltage
axis zoom ×2 to ×10; 3 settings, compression 1/2
Recorder function
Time axis
100 ms*
4
to 1 hr/division; 14 settings, 1 division = 80 samples,
time axis compression 1/2 to 1/50; 5 settings
*
4
100 ms to 500 ms/division ranges shown only on display when using AC Adapter. 100
ms to 1 s/division ranges shown only on display when using batteries
Sampling period
2.5 µs fixed
Recording length
20 to 400 divisions, "continuous"*
5
*
5
only “continuous” for X-Y plotting
X-Y sampling period
250 µs; fixed (dot), 500 µs to 10 ms (line)
X-Y axis resolution
20 pixel/DIV (display), 80 pixels/DIV × 80 pixels/DIV (optional printer)
Other functions
Back-scroll of memory data (max. last 400 divisions) and reprinting
of stored data
(with/ optional printer), logging (numerical printout) (with/
optional printer)
, voltage axis magnification ×2 to ×10; 3 settings,
compression 1/2; 1 setting. X-Y waveform plot
(one plot on the 8807-
01, up to three plots on the 8808-01)
RMS Recorder & Memory function (for 50/60 Hz and DC)
Time axis
RMS recorder: 100 ms to 1 hr/division; 14 settings
Memory recorder: 200 µs to 20 ms/division; 7 settings
1 division = 80 samples, time axis compression 1/2 to 1/50; 5 settings
Sampling period
RMS recorder: 250 µs fixed
(800 RMS data/second)
Memory recorder: 1/80 of time axis range
RMS calculation accuracy
±3% f.s.
Recording length
RMS recorder: 20 to 200 divisions, continuous
Memory recorder: 20 to 400 divisions, OFF
(only RMS recorder
when OFF)
Other functions
Back-scroll of memory data (max. last 200 divisions) and reprinting of
stored data
(with/ optional printer), for memory recorder: back-scroll of
memory data
(max. last 400 divisions) and reprinting of stored data (with/
optional printer)
, logging (numerical printout) (w/ optional printer), voltage
axis magnification ×2 to ×10; 3 settings, compression 1/2; 1 setting.
Auxiliary function
General
Printing of settings including input range, trigger time, etc., cursor
measurement, scaling, comment input, screen hard copy, start
condition retention, auto setup, auto saving, remote control, auto-
range setting, list & gauge printing
(with/ optional printer), DMM
function
(voltage shown as numerals on the display).
Calculation functions
(Memory recorder)
Up to four arithmetic operations simultaneously
Average value, effective
(RMS) value, peak to peak value, maximum
value, time to maximum value, minimum value, time to minimum value,
period, and frequency, area, X-Y area.
DMM function
Display update rate: 1 s, display contents: AC+DC rms
(measurement signal is DC, 50/60Hz only), or DC instantaneous value
Display digits: 4 digits
(last digit 0 to 4 is rounded zero, 5 to 9 is
rounded five)
Voltage range: Auto only (10 mV to 100 V/division, 5 settings)
Accuracy: ±3% rdg. ±5dgt.
Analog input
(accuracy at 23 ±5 ˚C/73 ±9 ˚F after 30 minutes warm-up time; accuracy guaranteed for 1 year)
Input Terminal: isolated BNC Inter-channel and input-frame isolation
Measurement
range
MEM or REC function: 10mV to 100V/DIV, 13 settings
RMS recorder function: 5mV to 50V/DIV, 13 settings
full-scale (f.s.) = 10 divisions, AC voltage for possible measurement
/ display using the memory function: 450 V AC rms, low-pass filter:
5/500 Hz, the measurement resolution is 1/160 of range
Maximum sampling rate
400 kS/s (simultaneous sampling of all channels)
Accuracy, frequency
characteristics
±0.5% f.s., DC to 50 kHz ±3 dB
Input resistance
and capacitance
1 MΩ, 7 pF approx. (at 100 kHz)
Input coupling
DC, GND
Max. allowable input
450 V AC rms, DC
(upper voltage which when applied to between input pins does not damage them)
Max. rated voltage
to earth
450 V AC, DC (upper voltage which when applied to input channel casing or
between input channels does not damage them)
Recording and display *
2
Waveform printing when the optional PRINTER UNIT 8992 is used.
Display method
5.7-inch STN color LCD, with Japanese/English selector
240 × 320 dots
*
2
Printer paper
112 mm (4.4 in) × 18 m (59.06 ft), thermal paper roll
*
2
Recording width
10 divisions in full scale, 1 division = 10mm (0.39in) (80pixels)
*
2
Paper feed density
8 rows/mm (203 rows/in)
16 rows/mm (406 rows/in) using the memory recorder's smooth print function.
*
2
Recording speed
Max. 10 mm/s (0.39 inch/s)
(when using AC Adapter),
max. 5 mm/s (0.2 inch/s)
(when using batteries)
Trigger function
Trigger source
Analog input CH1 - CH4
(8807-01: CH1 - CH2), logic input A - B,
external, timer, manual
(either ON or OFF for each source), logical
AND/OR of sources
Trigger types
(Analog)
Level: Triggered when set voltage value is exceeded in UP or
DOWN direction.
Window in/out: When entering or exiting a level range defined
by upper or lower limit
Voltage drop: Only for AC 50/60 Hz power lines. Triggered
when the peak voltage falls below setting value
RMS level: Only for DC and AC 50/60 Hz power lines.
Triggered when rms value crosses set value in UP or DOWN
direction
(RMS recorder function only)
Real-time waveform judgment: Only for AC 50/60 Hz power
lines. Trigger function that monitors when a signal exceeds the
evaluation area (Memory recorder function only)
Level setting
resolution
0.25% f.s., waveform judgment trigger only: 0.1% f.s. (f.s.=10
divisions)
Trigger types
(Logic)
Pattern trigger: 1, 0, or × (disregard), logical product (AND) or
logical sum (OR) set for 4 channels
Trigger filter
(analog/logic)
9 settings from 0.1 to 10.0 divisions or OFF
(memory recorder) ON/
OFF
(recorder)