Data Sheet
5
Convenient features such as the DMM function, special range
for a clamp probe, numerical value calculation, scaling, A/B
cursor measurement, free comment input, and automatic restart
after power outage make the measurement work quick and
simple.
Convenient features for ease of operation
Example showing measurement data imported to Excel.
■ Waveform Viewer software
Measurement data can be saved in binary format. Also,
can be converted to text format for numerical analysis in a
PC spreadsheet program.
Data can be saved in binary or text formats. The binary format is for
data to be used in the
MEMORY HiCORDERs 8807-01 and 8808-01.
Data saved to the PC in binary format can be converted to text format
using the supplied
Wv (Waveform Viewer program), for loading into a
spreadsheet program such as Excel.
■ Display copy in BMP format
Displayed images can be saved in BMP format to
easily create and print color reports from the PC’s word
processor.
Waveforms acquired by the memory recorder can be stored on
flash ATA-PC cards. Stored waveform data can be converted to
text (CSV) format files by the supplied Wv Waveform Viewer
PC application program.
Off-Line Data Exchange with a PC
- Data Communication with PC, other functions -
CONVERSION ADAPTER 9199
(receiving-end banana/BNC output)
Connects Indirectly
(Banana plugs)
■ DMM Function
Digital Multi Meter functions are provided
for simple input voltage checking. Selectable
modes are Effective value mode (AC+DC),
and Instantaneous value mode (DC), each
displaying four numeric digits. When the scaling
function is enabled, the specified scaling value is
incorporated.
Note: Convenient for checking waveform recordings of
power lines. RMS display is for 50/60 Hz or DC only.
Connects Directly
(BNC plug)
9018-50
CLAMP ON PROBE 9018-10
10 − 500 A f.s.
40 Hz − 3 kHz
Output 0.2 V f.s.
CLAMP ON PROBE 9132-10
20 − 1000 A f.s.
40 Hz − 1 kHz
Output 0.2 V f.s.
■
Special range for clamp probe enables easy current measurement
*
1
9018
■ RS-232C connection to PC
The PC and HiCORDER can be directly connected serially for transferring
recorded data and remote settings. The software created by the user may be
used on the PC.
Using the CLAMP ON PROBE 9018-10, current waveforms
can be captured on live lines. Voltage range settings and
scale settings are performed with a one-touch operation
thanks to the special clamp probe range provided.
*
1
Only compatible with the CLAMP ON PROBEs model 9018-10 and
9132-10.
The
CLAMP ON PROBEs model 9018 and 9132 can be connected
using the
CONVERSION ADAPTER 9199.