Chapter 3 Measurement Procedures
In DC A or AC+DC A mode, when OUT1 is set to MON,
output can be obtained through a low-pass filter (LPF) set
to approximately 1 Hz. This function will be useful for
outputting a DC component only when an AC component
or noise is superimposed on the DC component.
Press the MON.FL key (SHIFT
OUTPUT key) to
enable the function. Press it again to disable the function.
If this function is enabled in AC+DC A mode, OUT1
outputs the DC component while OUT2 outputs the AC
component (REC). The mode indicator changes from “
to “
“, and the reading shows the AC measurement. If
the function is disabled, the indicator returns to “
.”
3.14 MON Output Filter Function
When MON.FL is used, the response speed of
OUT1 will be reduced.
OUT1 has a 1-Hz filter. If the AC component
included in a waveform has is low frequency and
large input, the filter may be unable to com-
pletely remove the AC component.
Peak measurement using this function shows
the same result as peak measurement in AC A
mode (peak measurement of the waveform with
the DC component removed).