Data Sheet

2
Integral functions such as battery charging/discharging, integral AC current and repeat integral measurements are
fully supported. In addition, the "Operating Time/Rate" function shows the operating state of loads.
Comprehensive Current Integral Functions of the 3290-
10
Features
Polarity-specifi c integral function measures battery charging and discharging with one instrument
(positive, negative and total integral values)
Operating time/Rate display function measures load operating status
(with settable threshold value)
Polarity-specifi c peak current display function measures peak battery charge and discharge current with one instrument
Repeat measurement and data storage function lets you easily conduct the same measurement up to 20 times and then save that
data in the unit's memory.
Time-limited mean value display function shows the mean value within a specifi c integration time
(can be used in combination with
repeat measurement)
External DC power capability enables long-term measurements under battery power
Specifi cation 3290 3290
-10
Current
Measurement
Measurement modes DC, AC+DC, AC (RMS/MEAN) DC, AC+DC, AC (RMS)
Peak value display Displays absolute value (of waveform peak)
Displays polarities independently
(±peak value of waveform, in DC mode)
Maximum/minimum value display
Displays maximum, minimum and mean values Maximum and minimum values
(in AC and AC+DC modes)
AC fi lter function In AC and AC+DC modes, fi lter (fc = 550 Hz) ON/OFF
DC fi lter function
During DC and AC+DC waveform output, lter (fc = 1 Hz) ON/OFF
During DC waveform output, fi lter
(fc = 1 Hz)
ON/OFF
Separate AC/DC output function
Separate output of AC content and DC content
-
Output (connector 1) Current waveform (2-V range)/rms current (2-V DC range) switchable
Output (connector 2) Rms current/Low-battery warning switch Current integral value
(selectable +, -, total, 1-V DC/range)
Integral
Measurement
Timer settings
-
1 min. to 99 hours, 59 min
(for repeated measurement up to 20 times)
Elapsed time display Hours:minutes
(up to 2000 hours)
Mean value time limit display Mean value = integral value / integration time
Output (connector 1) Selectable current waveform or rms current
Output (connector 2) Current integral value
(selectable +, -, total, 1-V DC/range)
Rate
Measurement
Rate
-
Rate = operating time / total measurement time
Operating time measurement
Displayed operating time: Hours:minutes
(up to 2000 hours)
Frequency
Measurement
Range setting Auto/manual ranging Auto ranging
Output (connector 1) Frequency value
(1-V DC/range)
-
Output (connector 2) Rms Current
Data Storage
-
Peak, maximum, minimum, mean value, integral value, operating time,rate
Bar Graph Display
When DC center is zero, bar graph can display 10X magnification
Display of number of memorized data points
Display Refresh Rate
N: twice/second, F: 4 times/second
(3290-
10
is 10 times/second)
, S: once/3 seconds
* 3290-
10
DC mode is once/second
Measurement Response Time Switching F: 0.2 s, N: 0.8 s, S: 8 s switching
Power Supply
Battery and AC adapter Battery, AC adapter, external battery
(8.4 to 15.6 V DC)
3290/3290
-
10
Function Comparison
0.0 A
Operating Time
Current Threshold Setting
Operating
Time A
Non-Operating Time B
Operating
Time E
Non-Operating Time D
Total Measuring Time (A+B+C+D+E)
(Measurement may be repeated up to 20 times)
DC Load Current Measurement
+
-
Positive Integral
Negative Integral
Total Measuring Time: Z=(A+B+C+D+E)
Operating Time: A+C+E
Rate (%): (A+C+E)/Z × 100
+ Peak Value
- Peak Value
Total integral value = positive integral +
negative integral
Operating Time C
* Integration interval can be up to 2000 hours
DC Load Current MeasurementDC Load Current Measurement
Sensor Used
Current
Range
Integral Range (5 ranges) Measurement Range
9691/9692
20 A
10.000 Ah to 100.00 kAh
0 to ± 210.0 kAh (9691)
0 to ± 420.0 kAh (9692)
200 A
100.00 Ah to 1000.0 kAh
9693
200 A
100.00 Ah to 1000.0 kAh
0 to ± 4200 kAh
2000 A
1000.0 Ah to 10000 kAh
Current Integral Range Table
Convenient Repeat Measurement Function
Integral measurements during a specifi ed interval can be
repeated continuously up to 20 times.
Integral and numerical measurement data values
(maximum, minimum, mean and peak) plus operating
time and duty are stored in internal memory. Saved data
can be recalled and verifi ed at any time.