Enhanced Low-Frequency Current Measurement Analysis Capabilities
■ Measures current from 1 Hz, such as that used for inverter motor control.
Inverter motor current and rms waveforms are recorded simultaneously. Simultaneous recording of control signals is the best method
for evaluating and analyzing such devices during product development and maintenance.
Current
Recording
Data saved to PC Card at worksite.
After returning to the
office, data can be
analyzed and stored
on a PC.
* Recording basic waveform component with
3290 filter function active
* Waveform and rms values can be recorded simultaneously (MON & REC Mode)
Half-wave rectified waveform
during operation (MON)
Rms current recording (REC)
Recording Current Consumption of a Hair Dryer
PC Card-Compatible Recorder
Control Signal Recording
PC Card-Compatible Recorder
■
Comprehensive Basic Current Measurement Functions of the 3290
All functions needed for measuring DC and AC current and distortion waveforms in a single instrument
Optional Model 9400 CARRYING
CASE stores everything together
◆
Features
❍
AC+DC mode measures true rms current of full- and half-wave rectifi ed waveforms and inverter output waveforms from 1 Hz
❍
Peak mode measures the maximum waveform amplitude of in-rush current at device startup
❍
Maximum, Minimum and Mean value modes easily measure current variations in lines with severely fl uctuating loads
❍
Simultaneous waveform and rms output enables rms and waveform or frequency fl uctuations to be externally recorded together
❍
Selectable measurement response time enables setting the instrument response to suit changes in load current
❍
LPF (fc = 550 Hz) fi lters out unnecessary harmonic currents and noise
-3
-2
-1
0
1
2
3
-40 -20 0 20 40 60 80
9691
9692
9693
-4.0
-2.0
0.0
2.0
4.0
-40 -20 0 20 40 60 80
No.1
No.2
No.3
No.4
No.5
Example of temperature sensitivity
characteristics
-3
-2
-1
0
1
2
3
-40 -20 0 20 40 60 80
9691
9692
9693
-4.0
-2.0
0.0
2.0
4.0
-40 -20 0 20 40 60 80
No.1
No.2
No.3
No.4
No.5
Example of zero-point temperature
characteristic (Model 9693 SENSOR)
Clamp-On Current Sensor Temperature Characteristics
* The Hall-effect current detection element in each sensor has a specifi c zero-point temperature characteristic that
manifests as deviation (scatter)
0.0 A
Combined DC and AC waveform
0.0 A
Half-wave rectifi ed waveform
0.0 A
Full-wave rectifi ed waveform
Measures rms properly in the
AC+DC Mode
Deviation [%]
Deviation [%]
Temperature [°C] Temperature [°C]
Sensor
Used
Frequency
Range
Rated
Input
Measurement
Range
Max. Display
Output Terminal
Voltage
Normal
Measurement
Peak
Measurement
9691
DC to
10 kHz (-
3 dB)
100 A
rms
150 A
peak
20.00 A
25.00 A 50.0 A
100 mV/A
200.0 A
105.0 A 150.0 A
10 mV/A
9692
DC to
20 kHz (-
3 dB)
200 A
rms
300 A
peak
20.00 A
25.00 A 50.0 A
100 mV/A
200.0 A
210.0 A 300.0 A
10 mV/A
9693
DC to
15 kHz (-
3 dB)
2000 A
rms
2840 A
peak
200.0 A
250.0 A 500 A
10 mV/A
2000 A
2100 A 3000 A
1 mV/A
Current range table
(common to 3290 and 3290-