Data Sheet
■ Specifications
Option 9462 THERMISTER TEMPERATURE PROBE
●Measurable conductor diameter: 3282 ; ø46mm max. 3281
; ø33mm max. ●Operatable locations: up to 2000 m above sea
level ●Operatable temperature and humidity range: 0˚C to
40˚C, 80%rh or lower ●Auxiliary functions: recording (for V,
A, Hz measurements displays maximum value (MAX), minimum
value (MIN), and average value (AVE)), data hold (holds display),
auto power off (approx. 10 minutes, buzzer sounds just before
power is turned off, can be extended or cancelled)●Display
refresh rate: digital display FAST; approx. 4 times/second,
NORMAL; approx. 2 times/second , SLOW; 1 time/ 3 seconds, bar
graph display approx. 4 times/second ● Frequency
characteristic: 40Hz to 1kHz ● Effect of conductor
position: At any position based on the center of the core 3281 ;
within ±4.0%, 3282 ; within ±1.0% ● Effect of external
magnetic field: In an external magnetic field of 400 A AC/m
3281 ; 1.5A max. 3282 ; 0.2A max.●Circuit dynamic: when
performing full-scale input to the range 2.5 or less (600 A
[
3282 is
1000 A
]
, 600 V range is 1.7 or less)●Maximum useable circuit
voltage: 600 V rms (insulated conductor)● Standards
applying: Safety EN61010-1:2001 Pollution degree 2 over
voltage category IV (predicted transitory excess voltage 8000 V),
EN61010-2-031:2002, EN61010-2-032:1995 EMC EN61326:1997
+A1:1998+A2:2001 ●Voltage resistance: case-input terminal,
between clamp cores 6.68 kV AC for 15 seconds ● Power
supply: Layered magnesium battery (6F22) ✕1, 45 hours or less
(continuous use)● Dimensions /mass: 3281 ; Approx
62W✕ 218H✕ 39Dmm,350g 3282 ; Approx. 62W✕ 230H
✕39Dmm, 400 g
9462 THERMISTER TEMPERATURE PROBE
(optional)
● Measuring range:-50 to 150˚C/-58 to
302˚F● Measuring accuracy: -50 to 50
˚C±3˚C/50 to 100˚C±4˚C/100 to 150˚C±5˚C
Note: the temperature accuracy given above
is only the accuracy for the temperature
probe. ±2%rdg.±2˚C(˚F) is added for the actual accuracy of the 3281/3282
themselves.
Cord length
approx 1.2 m
The HAND STRAP not only prevents the
device from being dropped when working,
but also effectively imparts strength when
parting crowded wiring to clamp.
Designed for ease of use on the site !
3281
DIGITAL CLAMP ON HiTESTER
3282
DIGITAL CLAMP ON HiTESTER
(All include 9207-10 TEST LEAD, 9399 CARRYING CASE and HAND STRAP )
DANGER
1. In order to prevent short-circuits and injury, use the clamp product on electrical
circuits with a voltage less than the maximum operation circuit voltage.
2. In order to prevent short-circuits and injury when the clamp core tip is open, do
not use bare conductors.
!
Inspect the unit and check that it is operating correctly before use. When carrying out
measurement on live lines, wear proper protective gear, insulating rubber gloves, insulating rubber
boots and safety helmet, and use extreme caution to avoid electric shock accidents.
WARNING
The HAND STRAP fits the device neatly to the hand
!
Function Mode Range
Accuracy
Maximum
permissible
input
AC current
A
RMS
Effective
value
30.00 ±1.0%rdg.±0.7%f.s.
(40 to 1 kHz)
600A rms continuous
1700A max.
300.0 ±1.0%rdg.±5dgt.
*1
(45 to 66 Hz)
1000
[600]
±1.0%rdg.±5dgt.
*1
(45 to 66 Hz)
[±1.0%rdg.±5dgt.
*1
(45 to 66 Hz)
]
1000A rms 5 minutes
[
600A rms continuous
1000A max.
]
PEAK
Peak value
30.0 ±5.0%rdg.±5dgt.
600A rms continuous
1700A max.
300 ±3.0%rdg.±5dgt.
1000
[600]
±3.0%rdg.±5dgt.
[±3.0%rdg.±5dgt.]
1000A rms 5 minutes
[
600A rms continuous
1000A max.
]
AC voltage
V
RMS
300.0/600
±1.0%rdg.±3dgt.
*2
(45 to 66 Hz)
600V rms continuous
1000V max.
PEAK
300/600 ±3.0%rdg.±5dgt.
Crest factor
1.00 to 5.00
±10.0%rdg.±5dgt.
Effective in the voltage
and current functions
Frequency
Hz
100.0
1000
±0.3%rdg.±1dgt.
(30 to 99.9 Hz)
±1.0%rdg.±1dgt.
(95 to 1000 Hz)
Resistance
Ω
1k/10.00k
±1.5%rdg.±5dgt.
Overload protection :
600V rms
Open terminal voltage
:
3.0V DC max
.
(resistance,continuity
functions)
Temperature
probe sold separately
˚C
-50 to 150˚C
±2.0%rdg.±2dgt.
Add the accuracy of the 9462
˚F
-58 to 302˚F
continuity
1kΩ
Buzzer sounds at
30 Ω ±5 Ω or less
*1. 40 to 45Hz , 66 to 1kHz ±1.5%rdg.±5dgt. *2. 40 to 45Hz , 66 to 1kHz ±1.5%rdg.±3dgt.
Measurement accuracy applies to input of at least 10% of the current, voltage and resistance range.
Display
LCD, digital (3000 counts, 999 counts (peak)
), bar graph (35 segments)
Even if voltage is mistakenly
applied to the resistance range,
the internal excessive voltage
protection element (PTC
thermister) protects the circuits
for up to 600 V AC.
NOT
FUSE
NEEDED
●Non-fuse type protects up to 600V AC
●Auto-power off to conserve battery life
Prevents careless mistakes
[ ] in the Specification :3281 23˚C±5˚C, 80%
or less.
The photograph is 3281.
9207-10 TEST LEAD
Lead length
approx. 70 cm
Accessories
9399 CARRYING CASE
CAT IV 600V
3281E5-4ZM-02U Printed in Japan
All information correct as of Dec.17, 2004. All specifications are subject to change without notice.
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
DISTRIBUTED BY
Shanghai Representative Office :
1704 Shanghai Times Square Office
93 Huaihai Zhong Road
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090, 0092
FAX +86-21-6391-0360
E-mail: info@hioki.cn