Data Sheet
12
■ Accessories
3196 POWER QUALITY ANALYZER
(9438-02 VOLTAGE MEASUREMENT CABLE (one each of red, yellow, blue
and gray, plus four black lines, Cord length: 3 m (9.84 ft), 9459 BATTERY PACK,
9458 AC ADAPTER, Strap, LAN connector cover, Input Cord Label, Operating
Manual (CD-R), Quick Start Manual)
By itself, the 3196 is only capable of voltage measurement. Purchase the optional
9660 or 9661 CLAMP-ON SENSOR for current and power measurement.
● Standard combination example
For three-phase 3-wire (3P3W3M) and three-phase 4-wire measurements
Models 3196 + 9661 (500 A) × 3 + 9339 + PC card (128 MB)
9670 PRINTER option components
The 9671 AC ADAPTER should be purchased along with the 9670 PRINTER.
Also, the 9638 RS-232C CONNECTION CABLE or RS-232C cable (9- to
25-pin crossover) is required to connect to the 3196.
A battery pack and battery charger to power the
9670 Printer are also available in
some countries. Please contact your HIOKI distributor for details.
7.2 V DC, 2700 mAh
100 to 240 V AC, 1.2 A
50/60 Hz
■ Option Specifications
9458 AC ADAPTER
Clamp On Sensors
9694
Primary current rating
5A AC
9661 9669
Accuracy Amplitude
(45 to 66 Hz) Phase
±0.3% rdg. ±0.02% f.s. ±0.3% rdg. ±0.02% f.s.
Appearance
±2˚ or less ±1˚ or less (±1.3˚ for 90 A or more)
Frequency characteristic
±1.0% or less for 66 Hz to 5 kHz (deviation from specified accuracy)
Effect of external magnetic field
Corresponding to 0.1 A or less (with magnetic field of 400 A/m AC)
Effect of conductor position
±0.5% or less
300 Vrms (insulated conductor)
Maximum rated voltage to earth
300 Vrms (insulated conductor)
Maximum allowable input (45 to 66 Hz)
50 A continuous 130 A continuous
Measurable conductor diameter
φ15 mm (0.59") or less φ15 mm (0.59") or less
Dimensions and weight
46W (1.81") × 135H (5.31") ×
21D (0.83") mm, 230 g (8.1 oz.)
46W (1.81") × 135H (5.31") ×
21D (0.83") mm, 230 g (8.1 oz.)
±0.3% rdg. ±0.01% f.s. ±1.0% rdg. ±0.01% f.s.
±0.5˚ or less ±1˚ or less
±2.0% or less for 66 Hz to 5 kHz (deviation from specified accuracy)
600 Vrms (insulated conductor)
550 A continuous
Corresponding to 1 A or less (with magnetic field of 400 A/m AC)
φ46 mm (1.81") or less
±1.5% or less
77W (3.03") × 151H (5.94") ×
42D (1.65") mm, 360g (12.7 oz.)
600 Vrms (insulated conductor)
1000 A continuous
φ55 mm (2.17") or less, 80 (3.15") × 20 (0.79") mm busbar
99.5W (3.92") × 188H (7.40") × 42D (1.65") mm,
590g (20.8 oz.)
Output voltage
10 mV/A AC 1 mV/A AC 1 mV/A AC 0.5 mV/A AC
9667
Appearance
Primary current rating
500 A AC, 5000 A
Output voltage
500 mV AC f.s.
Accuracy Amplitude
(45 to 66 Hz) Phase
±2.0% rdg. ±1.5 mV
(for input 10% or more of the range)
±1˚ or less
Frequency characteristic
±3 dB or less for 10 Hz to 20 kHz (deviation from specified accuracy)
Effect of external magnetic field
Corresponding to 5 A, 7.5 A max. (with magnetic field of 400 A/m AC)
Effect of conductor position
±3.0% or less
Maximum rated voltage to earth
1000 Vrms (insulated conductor)
Maximum allowable input (45 to 66 Hz)
10000 A continuous
Measurable conductor diameter
φ254 mm (10") or less
Dimensions and weight
Sensor length: 910 mm (2.99 ft), 240 g (8.5 oz.),
Circuit: 57W (2.24")
× 86H (3.39") × 30D (1.18") mm, 140 g (4.9 oz.)
9290-10 CLAMP-ON ADAPTER
Cord length : 3 m (9.84 ft)
Up to 1500 A AC, CT ratio : 10:1
Measurable conductor diameter :
φ55 mm (2.17"), width : 80 mm (2.17") bus bar
Standard accessories
9660
Cord length: 3 m (9.84 ft)
Cord length: 3 m (9.84 ft)
CAT III 300V CAT III 300V CAT III 600V
CAT III 1000V
Cord length
Sensor to circuit: 2 m (6.56 ft)
Circuit to connector: 1 m (3.28 ft)
Cord length: 3 m (9.84 ft)
■ Options
9660 CLAMP ON SENSOR (100 A AC) Voltage output type
9661 CLAMP ON SENSOR (500 A AC) Voltage output type
9667
FLEXIBLE CLAMP ON SENSOR (5000 A AC) Voltage output type
9445-02 AC ADAPTER (for the 9667, for America, Japan)
9445-03 AC ADAPTER (for the 9667, for Europe)
9669 CLAMP ON SENSOR (1000 A AC) Voltage output type
9694 CLAMP ON SENSOR (5 A AC) Voltage output type
9657-10
CLAMP ON LEAK SENSOR (5A AC with Model 3196) Voltage Output Type
9675
CLAMP ON LEAK SENSOR (5A AC with Model 3196) Voltage Output Type
9290-10 CLAMP ON ADAPTER
9264-01 WIRING ADAPTER (3P3W)
9264-02 WIRING ADAPTER (3P4W)
9438-02 VOLTAGE MEASUREMENT CABLE
(standard accessory)
9459 BATTERY PACK (standard accessory)
9670 PRINTER (with one roll recording paper)
9671 AC ADAPTER (for 9670)
9237 RECORDING PAPER
(80 mm (3.15") x 25 m (82.03 ft), 4 rolls, for 9670)
9638 RS-232C CABLE (1.5 m (4.92 ft), for printer connection)
9642 LAN CABLE (5m (16.41 ft), with straight and crossover connectors)
9339 CARRYING CASE (soft)
9340 CARRYING CASE (hard)
9624 PQA-HiVIEW (PC application software)
9624-10 PQA-HiVIEW PRO (PC application software)
*Please select either USB or Parallel license Key.
9726 PC CARD 128 M
9727 PC CARD 256 M
9728 PC CARD 512 M
Operating Manual (bound version)
XD112 GPS Box (including antenna and RS-232C cable)
CAT III 600V
104W (4.09") × 51H (2.01") ×
18D (0.71") mm, 250 g (8.83 oz.)
9339 CARRYING CASE
(soft case)
9340 CARRYING CASE
(hard case)
450W (17.72") × 350H (13.78") × 200D
(7.87") mm, 3.0 kg (106.01 oz.)
380W (14.96") × 560H (22.05") × 260D
(10.24") mm, 6.3 kg (222.2 oz.)
9459 BATTERY PACK
Power supply
LR03 alkaline battery × 4
(continuous operation max. 168 hours)
or
9445 AC ADAPTER
(optional)
Clamp On Sensor
Cord length: 3 m (9.84 ft)
100 A AC 500 A AC 1000 A AC
3196E7-61B-05P Printed in Japan
All information correct as of Jan. 6, 2006. All specifications are subject to change without notice.
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
DISTRIBUTED BY
Shanghai Representative Office :
1704 Shanghai Times Square Office
93 Huaihai Zhong Road
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090, 0092
FAX +86-21-6391-0360
E-mail: info@hioki.cn
100 to 240 VAC, 50/60 Hz
134W (5.28") × 70H (2.76") × 41D (1.61") mm
350 g (12.37 oz.)
9671 AC ADAPTER