Guidance and manufacture’s declaration – electromagnetic immunity
TheDBP-01P DBP-01HP is intended for use in the electromagnetic environment specified
below. The customer or the user of theDBP-01P DBP-01HP should assure that it is used in
such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted
RF
IEC
61000-4-6
3 V
rms
150 kHz to 80
MHz
Not
applicable
Portable and mobile RF communications
equipment should be used no closer to any part of
theDBP-01P DBP-01HP, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to 2.5
GHz
3 V/m