深圳星脉医疗仪器有限公司 upper arm blood pressure monitor User manual (Apply to:Q06、Q06B) Edit:2020-11-22 Version: A/0 深圳星脉医疗仪器有限公司
深圳星脉医疗仪器有限公司 About this manual The manual mainly introduces the installation and application method of upper arm blood pressure monitor.Users should read carefully before application (include warnings, contraindications and notes). Manufacturer Shenzhen Hingmed Medical Instrument Co., Ltd. Manufacturing address: 4th Floor, Zhonghangfeixiang Building, NO.
深圳星脉医疗仪器有限公司 Table of Content 1.Introduction ................................................................................................................................................ 4 1.1product feature ................................................................................................................................... 4 1.2Safety Instructions.............................................................................................................................. 4 1.
深圳星脉医疗仪器有限公司 1.Introduction 1.1product feature Thank you for purchasing the HINGMED upper arm blood pressure monitor.This monitor can be used by multiple users,This blood Pressure monitor use the oscillometric method of blood pressure measurement.It has some functions that measure the blood pressure and display the result.When measuring the blood pressure,the user need to wear the cuff on the upper arm. 1.
深圳星脉医疗仪器有限公司 Model Q06B /Q06 Measuring method Oscillometric method Range From 0 to 289mmHg Pulse range 40-200bpm Resolution ratio 1mmHg Accuracy ±3mmHg (pulse rate ±3bpm or ±3% whichever is grater) Power DC 3.7V Li-ion Cuff size 22cm-41cm Size or 22cm-36cm L:116*W:58*H:17.
深圳星脉医疗仪器有限公司 *7.9.2.14please use the adapter which output DC 5V 2A to charge the device When the device start inflating may have a little uncomfortable. Do not leave the device unattended with infants or person who can express their consent. Do not apply strong shocks and vibrations to or drop the device and arm cuff. Do not take measurements after bathing,drinking alcohol,smoking,exercising or eating. Do not wash the arm cuff or immerse it in water.
深圳星脉医疗仪器有限公司 * In case of equipment failure, the error code will be displayed in the format of "EC XX", see the section on Common Failures for details. * State that the blood pressure measured should be explained by a professional. The product is not intended for newborns.
深圳星脉医疗仪器有限公司 *Three hundreds of discharge cycles after which a rechargeable battery needs to be replaced * Please put the device keep away the children and pet The list of device that can potentially cause interference problem,eg. Microwave oven; 3. Product structure and operation principle 3.1The component of product No. 1 2 3 4 Compment Main body USB wire Cuff User Manual Count 1pcs 1pcs 1pcs 1pcs 3.2Name and description of each part size 1M 17-41cm or 17-36cm A D E B C NO.
深圳星脉医疗仪器有限公司 CF type applied part Series number Manufacturer Valid date Keep dry This way up Fragile Complying standard Stacking number limit by with WEEE Temperature limitation Refer to user manual Date of manufacture General warning sign General Prohibition sign 2:protect against solid foreign General Mandatory action sign Protected against solid foreign objects of 1 2,5 mm IP22 and greater 2:Protection against vertically falling water drops when Enclosure tilted up to 15° Atmospheri
深圳星脉医疗仪器有限公司 (1)press the "start/stop" button,the cuff start to inflate automatically,increasing number on the display ,remain still and do not move your arm and until the measurement process is complete. (2) When the measurement is complete,your cuff completely defeat, your blood pressure and pulse rate are displayed and broadcast. 4.
深圳星脉医疗仪器有限公司 EC09 The measurement timeout, in adult mode, Wait exceeded 120 seconds EC 11 EC26 If 2-3minutes Take another measurement Push the “start/stop” button to Stop Push the “start/stop”button to take another measuring measurement Battery is worn Please recharge there is other error code ,please contact customer service.
深圳星脉医疗仪器有限公司 6.maintenance ************************************************************ Note: * Sphygmomanometers and accessories need not be sterilized, but should be kept clean. If there is pollution, it should be promptly cleaned before disinfection. *Suggestion that clean the device by every month; To protect the device from damage,please observe the following: Do not subject the device to extreme temperature,humidity,moisture or direct sunlight. Do not wash the arm cuff or immerse it in water.
深圳星脉医疗仪器有限公司 8、Component replacement list Caution:Please use the component supply by the manufacture ,otherwise might cause the measurement error or injury NO. Component Type 1 Battery 1020mAh,3.
深圳星脉医疗仪器有限公司 9、 Warranty Card Warranty Card Product model and SN code: Name: Purchase Date: Address: Dealer: Tel: Postal Code: Dealer stamp: The limited liability of guarantee Hingmed provides the original purchaser the following limited warranty from the date of invoicing. upper arm blood pressure monitor………………………………………… 12 months Hingmed warrants each monitor to be free from defects in material and workmanship.
深圳星脉医疗仪器有限公司 10 EMC information This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment. Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
深圳星脉医疗仪器有限公司 Guidance and manufacture’s declaration – electromagnetic immunity TheDBP-01P DBP-01HP is intended for use in the electromagnetic environment specified below. The customer or the user ofDBP-01P DBP-01HP should assure that it is used in such an environment. Immunity test Electrostatic discharge (ESD) IEC 60601 test level Compliance level ±8 kV contact ±8 kV contact ±15 kV air ±15 kV air Electromagnetic environment guidance Floors should be wood, concrete or ceramic tile.
深圳星脉医疗仪器有限公司 environment. NOTE UT is the a.c. mains voltage prior to application of the test level. Guidance and manufacture’s declaration – electromagnetic immunity TheDBP-01P DBP-01HP is intended for use in the electromagnetic environment specified below. The customer or the user of theDBP-01P DBP-01HP should assure that it is used in such an environment.
深圳星脉医疗仪器有限公司 Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
深圳星脉医疗仪器有限公司 Separation distance according to frequency of transmitter (m) Rated maximum output power of transmitter 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz 0.12 0.12 0.23 0.38 0.38 0.73 1.2 1.2 2.3 3.8 3.8 7.3 12 12 23 (W) 0.01 0.