User Manual
3. Technical Data
Parameter
Value
Temperature Range
Adjustable: environment temperature+5 °C (9 °F), up
to 50°C (122 °F)
Effective Radiating Area
Ф 70 mm (2.76”)
Temperature Resolution
0.1 °C (0.18 °F)
Temperature Accuracy
± 0.1 °C (0.18 °F)
Temperature Stability
± 0.1 °C/h
Effective Emissivity
0.97 ± 0.02
Power Consumption
20 W
Environment Temperature
0 °C to 35°C (32 °F to 95 °F)
Weight
1.85 Kg (4.08 lb)
Dimension
120 mm × 176.6 mm × 119.6 mm (4.72” × 6.95” × 4.71”)
Power
100 VAC to 240 VAC
Mounting
Mounting on a plane desk
or a tripod with 1/4-20UNC Interface
or a ceiling mount bracket with 1/4-20UNC Interface
4. Mounting
1. Check the power supply. The working voltage should be 100 VAC-240 VAC, 35 W.
2. Check if the desk is horizontal.
3. Mount the device on the desk or a tripod in front of the thermal camera firmly. Make
sure that the front of black body can show in the live view of thermal camera
completely, and outside the temperature measurement area.
4. Switch on the device, and the power indicator is on. The factory default setting
temperature is 40 °C.
5. Operation
1. Connect the black body with a 10 A socket via the power cable.
2. Switch on the device, and the power indicator is on. The factory default setting
temperature is 40 °C.
3. Wait for 30 minutes after switching on device. You can calibrate the thermal camera
when the temperature is stable at 40 °C.
4. Switch off the black body and store it after use.
6. Maintenance
1. The black body should be managed and maintained by professional person, and
maintenance and usage should be recorded.
2. DO NOT use the device in an environment with oil or chemical air.
3. To ensure the temperature accuracy, place the device firmly in the environment with
stable temperature. The recommended using environment temperature is 10 °C to
25 °C (50 °F to 77 °F)
4. Power off the device before mounting or removing it.
5. Clean the surface of device with a neutral detergent. Clean the dust on the black body
with a soft brush.
6. Store the device in the packing box after use to prevent dust. Make sure the
environment temperature is suitable.
7. DO NOT dismantle the device without authorization. Contact the professional
specialist for repairing.
8. To ensure the temperature accuracy, it is recommended to calibrate the black body
once a year. Please send the device to qualified calibration institute.
UD22526B