Chapter 4 Temperature Measurement
The thermography (temperature measurement) funcon provides the real-me temperature of
the scene and display it on the le of your screen. The thermography funcon is turned on by
default.
4.1 Set Thermography Parameters
You can set thermography parameters to improve the accuracy of temperature measurement.
Steps
1. Go to Local Sengs → Thermography Sengs .
2. Set the Thermography Range, and Emissivity, etc.
Thermography Range
Select the temperature measurement range. The device can detect the temperature and
switch thermography range automacally in Auto Switch mode.
Emissivity
Refer to Common Material Emissivity Reference to set the emissivity of your target.
Reecon Temperature
If any object (not the target) of high temperature is in the scene, and the target emissivity is
low, set the reecon temperature as the high temperature to correct the thermography
eect.
Distance
The distance between the target and the device. You can customize the target distance or
select the target distance as Near, Middle, or Far.
Environment Temperature
The average temperature of the environment.
Humidity
Set the
relave humidity of current environment.
Opcal Transmissivity
Set the opcal transmissivity of external opcal material (e.g.: germanium window) to
improve the temperature measuring accuracy.
External Opcs Temperature
Set temperature of the external opcal material (e.g.: germanium window).
3. Return to previous menu to save the sengs.
Handheld Thermography Camera User Manual