Chapter 4 Temperature Measurement
The thermography (temperature measurement) funcon provides the real-me temperature of
the scene and display it on the le of your screen. The thermography funcon is turned on by
default.
4.1 Set Thermography Parameters
You can set thermography parameters to improve the accuracy of temperature measurement.
Steps
1. Go to Local Sengs → Thermography Sengs .
2. Set the Thermography Range, and Emissivity, etc.
Thermography Range
Select the temperature measurement range. The device can detect the temperature and
switch thermography range automacally in Auto Switch mode.
Emissivity
Refer to Common Material Emissivity Reference to set the emissivity of your target.
Reecon Temperature
If any object (not the target) of high temperature is in the scene, and the target emissivity is
low, set the reecon temperature as the high temperature to correct the thermography
eect.
Distance
The distance between the target and the device. You can customize the target distance or
select the target distance as Near, Middle, or Far.
Environment Temperature
The average temperature of the environment.
Humidity
Set the
relave humidity of current environment.
Opcal Transmissivity
Set the opcal transmissivity of external opcal material (e.g.: germanium window) to
improve the temperature measuring accuracy.
External Opcs Temperature
Set temperature of the external opcal material (e.g.: germanium window).
3. Return to previous menu to save the sengs.