User's Manual
11
instrument) , “RB’(reflector board);
5) PSM: Prism constant .Generally as “0” or “30”,if the prism is
special, you need to input itself constant value;
6) Mode: Distance measurement mode .You have four options to
select, which are “Single”, “Rept.”, “Avg.”(Set by “count”),
“Track”(Fast but low accuracy) . The “Single” measurement
and “Avg.” measurement can be ended automatically after a
successful measuring. But if you want to end the “Track” and
“Rept.” measurement, you must press the key [ESC];
7) Count: Times of the “Rept.” measurement. The range is “1” to
“9”;
8) Record: to set the mode of storing data under the basic
measurement. You can choose “ENT” (press [ENT] to enter
the “rec. data” interface.), “AUTO” (record the measured
data automatically after a successful measurement),
“NO”(won’t record the measured data , even though pressing
[ENT]).
3.4 Settings of hot key[•]
Under the basic measurement interface, you can press the
hot key [▪] to enter the “Hotkey” interface.
----Hotkey----
1.T.H
3.Note
2.Temp-Press
3.4.1 Input of target height
If you want to change the current system default of target