User's Manual

100
----Calibrate----
1. Adjust I.E
2. Input INS. const.
3. Adjust X
4. Adjust Y
Take positive
ESC ENT
Info
2) Aim at the target in face left, then press [ENT];
3) The instrument measures the vertical angle
automatically with a prompt displaying angle , then you
can press [ENT];
4) A prompt Take reverse appears, you can aim at the
target in face right, then press [ENT];
5) The instrument measures the vertical angle and
calculates and displays the index error automatically
to ask you whether want to save the result ,you can
press [ENT] to save and exit the interface(the index
error completed) ,or press [ESC] not to save (the index
error keep the original value.).
11.6.2 Input the instrument constant
1) In the calibration menu, press [2] or press [ENT] after
moving the cursor to Input INS. const. to enter the
interface of Config ,as shown in picture below:
Config
Add C:
Mul C:
SAVE
5
0
mm
1
mm
2) Input parameter;
[Notice*]:
The range of additive constant (ADD C) is
-99~+999mm;