User's Manual

99
4) When the cursor is moved to Offset-> or
Program->, you can press [ENT], then a prompt of
sub-functions will appear and display, as shown in
picture below;
User keys 1
BS reset.
Offset->
Temp. & Press
Program->
File Manager
Offset
Menu
Dist
Angle
2D
User keys 1
BS reset.
Offset->
Temp. & Press
Program->
File Manager
Prog./Opti
Proj.[Menu]
Project
Arc
MLM(AB-AC
5) Press [ENT] to select the function to define the user
key and return to the User keys menu.
11.6 Calibration
This function is used to calibrate the measurement
parameter of instrument.
In the basic measurement interface, press [ENT] to enter
the interface of menu, then press [6] or press [ENT] after
moving the cursor to Calibration to enter the interface of
Calibration, as shown in picture below:
----Calibration----
1. Ajust I.E
3. Adjust X
2. Input INS. const.
4. Adjust Y
11.6.1 Adjusting index error(I.E)
1) After enter the interface of calibration menu , press [1]
or press [ENT] after moving the cursor to Adjust I.E
with a prompt Take positive appearing, as shown in
picture below;