Express 3 Primary Tube Centrifuge Operator’'s Manual Sl Procession A Division of IRIS International, Inc.
Operator’s Manual Express 3 Primary Tube Centrifuge Model Number M502-22 For in-nitro diagnostic use Product Number S$SX3 Express 3 for 100-240 VAC, 50/60 Hz Stat Spin and Iris are registered trademarks of IRIS International, Inc. Copyright 2006 Printed in U.S.A.
Table of Contents How to use this manual.. Unpacking and Installation .. Inspect Packaging . Verify Contents Install Connect Power System Overview . Principle and Intended Use . Symbols and Definitions . Operator Controls .. Error Indicators .. Accessories. Section Operating Instructions Opening and Closing the Cover . Removing and Installing the Rotor . Spinning the Sample . Maintenance .. Overview .. Cleaning .. Checking the Rotor Speed. Troubleshooting . .9 Service . 10 Section 5 .
How to use this manual This manual along with information contained on product labels and in package inserts should provide you with all the information you need to operate and maintain the Express 3. Notes appear in italics to highlight information. ‘When the Information requires special attention, a caution symbol A appears next to the Italicized text.
Section 1 Unpacking and Installation Inspect Packaging The Express 3 and s accessories are delivered In a single carton. If the centrifuge or accessories have been damaged In transport, please Inform your carer Immediately. NOTE: Save ihe shipping carton and packing materials to facilitate erratum should service be required.
Section 2 System Overview Principle and Intended Use VD For in-nitro diagnostic use to produce the rapid separation of whole blood contained in original collection tubes. The Express 3 is a high-speed bench top centrifuge used to rapidly separate blood components in the original sample collection tubes. The sight-place 30° angle rotor included with the centrifuge produces rapid separation while minimizing potential for re-suspension of cells. The operator selects a fixed spin time minutes.
Operator Controls > Tha Stop/Open: button interrupts the cycle and Stop/Open releases the: cover when the rotor has stopped: © 2 min © min The 3 min button selects a.cycle of three minutes at Moy ransom. Servicewoman. Pashas to signet an error condition or remains continuously illuminated when service Is needed: The red:indicator with an adjacent wrench:symbol @ Error Indicators Express 3 error conditions are signaled by a flashing red Error/Service indicator on the control panel. tors.
Section 3 Operating Instructions Opening and Closing the Cover An electronic operated cover Interlock mechanism prevents operation If the cover Is not completely closed and locked, and prevents the cover from opening when the rotor Is muting. An operating cycle can be initiated only if the cover latch is engaged. Cover Interlock By-pass In case of power allure or malfunction, the Stop/Open button may not release the cover.
2. Slide the rotor onto the shaft and push down gently to seat the drive pins in the holes. 3. Place the retaining nut on the shaft and tighten securely using the 5/16 inch hex key. You will not be able to tighten the retaining nut If the drive pins are not properly aligned. if this occurs Sift the rotor off and repeat the process until it is seated. Spinning the Sample & CAUTION Never operate the Express 3 without the rotor docked securely in place.
Unloading 1. Upon completion of the cycle, the rotor decelerates to a complete stop in 30 seconds or less, a signal beep will sound, and the cover latch automatically releases. Lift the cover. 2. 3. Carefully remove the tubes using care not to re-suspend the sample.
Section 4 Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenance on all instruments. Contact customer service department or your distributor if the instrument is not functioning property. CAUTION Unplug the Express 3 from the wall outlet before cleaning or performing maintenance.
Troubleshooting No power light N Is the line cord complacently plugged Into lhf wall? Yes + Is the wall outlet working properly? + Yes v Is the fuse blown? v No ¥ Contact Iris Sample Processing or an Authorized Repair Facility. — No—> — No—> Yes—> Plug Lina cord into the wall. Try another wall outlet. Replace fuse. Centrifuge will not spin, or shuts off prematurely v Is the cover closed and latched? + Yes ¥ |s the error light flashing? v No + Contact Ils Sample Processing or an Authorized Repair Facility.
Service Refer all service to Iris Sample Processing or an Authorized Repair Facility. 1-800-782-8774 Instrument fuse(s) are located on the rear panel. Remove the line cord to access the fuse holder. Replace with fuses of the same types and rating. 30A250VF 5x20 mm Be sure to complete the Warranty Registration Card as directed or go to compromise/ to obtain a copy of the warranty card.
Section 5 Appendix Specifications Product Number 88X3 Span Parameters 2 min: 120 seconds @ 7200 rpm /4440 xg 3'min; 180 seconds’ @ 7200 rpm /4440 xg 5.min: 300 seconds @ 5600 pm £2685 xg Deceleration Time = 30 seconds: Dimensions Height: 77 Width: 117 Depth: 10 Weight: 8% 1bs. Environmental For Indoor Use onl Specifications are subject to change without notice. minimum operating temperature.
References 1. CLII. “Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline-Second Edition.” CLII document M29-A2 [ISBN 1-56238-453-8]. CLII, 940 West Valley Rd, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2001. 2. CDC. Recommendations for Prevention of HIV Transmission in Health Care Settings. MW (Suppl. No. 1987. 3. CDC.
Iris Sample Processing Warranty Iris Sample Processing, a Division of IRIS International, Inc. warrants that the instruments shale be free from defects in material and/or workmanship, under normal use and service, for the period expiring twenty-four (24} months from the date of installation, provided the purchaser has completed and forwarded to Iris Sample Processing the Warranty Registration Card.