Product Manual

Technical Specifications
Principle Modified glucose dehydrogenase in
which the total amount of glucose
is measured at the end point
photometrically
Calibration Factory calibrated and traceable to the
ID GC-MS method; needs no further
calibration and no coding
Sample Material Capillary, venous or arterial whole blood
Measurement
Range
Plasma equivalent values:
0–24.6 mmol/L (0–444 mg/dL)
Whole blood values:
0–22.2 mmol/L (0–400 mg/dL)
Results Including data entry, within one minute
for normal glucose levels
Sample Volume < 5 µL
Dimensions Analyzer: 170  ×  93  ×  50 mm
(6.70  ×  3.66×  1.97 inches)
Docking Station: 206  ×  135  ×  61 mm
(8.10  ×  5.30  ×   2.40 inches)
Weight Analyzer: 350 g (0.77 pounds) with
batteries installed
Docking Station: 566 g (1.24 pounds)
Storage Temp. Analyzer: 0–50 °C (32–122 °F)
Microcuvettes: below 8 °C (46 °F),
room temperature for up to 3 days;
one month open vial stability.
Operating Temp. 1530 °C (59–86 °F)
Power Options Internal rechargeable Li-ion batteries or
docking station with AC adapter
Interface USB/LAN POCT1-A
Quality Control Built-in ”selftest”; system can be
verified using liquid controls
Vista 1500 (mmol/L)
HemoCue 201 DM (mmol/L)
0
0
5 10 15 20 25
5
10
15
20
25
HemoCue Glucose 201 DM vs Vista 1500, individual
replicates. No of replicates = 44. Origin from
unpublished data ref Dr. S. Kos, Dr. E. Eppens,
A. van Meerkerk, MaastadLab, Maasstad Hospital,
Rotterdam, The Netherlands
Comparison of the HemoCue 201 DM System
with Vista 1500 (Siemens)
y = 1,010x - 0,187
r= 0,990