User Manual

30 MMS
®
Inspection DFT High
Calibration
Enabling the recalibration function Quick
Enable the Quick function for the desired calibration: Main Menu () >
Calibrations > Name of calibration > Quick > OK ( = Function is
enabled)
Disabling the recalibration function Quick
Disable the Quick function for the desired calibration: Main Menu () >
Calibrations > Name of calibration > Quick > OK ( = Function is
disabled)
7.3.4 Calibration step Zero
Measurement on the uncoated reference part
Required material
Base material: Magnetizable material = Fe: Ferromagnetic reference part
without the coating to be measured from customer's own production.
Gage type FE and FE + NF, use of the magnetic induction test method,
NF/FE display screen
Base material NF: Non-magnetizable, electrically conducting material =
NF = Non-magnetic, electrically conducting reference part without the
coating to be measured from customer's own production. Gage type
FE + NF, use of the amplitude-sensitive eddy current test method.,
NC/NF display screen