User Manual

20 MMS
®
Inspection DPM High
Measurement
6.1 Before you start
The gage is switched on ( key)
Temperature probe
The gage surface temperature probe is used: The protective cap has
been removed
A magnetic surface temperature probe is used (example): The
temperature probe is connected to the gage
The required batch is opened ( key in the main menu, see also
page 18)
6.2 Measurement procedure
Air temperature (Ta), measured continuously
Relative humidity (%RH), measured continuously.
Surface temperature (Ts)
Measuring with the gage surface
temperature probe: Press the surface
probe on the surface until the value for Ts
hardly changes any more.
Alternatively, you can measure the
surface temperature by connecting a
Type K surface probe to the gage
To manually save all climate factors: >
To continuously log all climate factors (log func-
tion): >
The top left corner of the window shows the
remaining time in minutes up to the next save
(e.g.: save interval 3600 min (= 6 h), every 3600 minutes, the climate
factors measured and calculated at this point in time are saved;
remaining time to the next save is 3598 minutes and 45 seconds)
Stop continuous logging (log function):
Gage
surface tem-
perature
probe
Surface
Spring-loaded temperature sensors
3598:45
Batch 001
%RH 34.9
Ta 26.1 °C
A