User Manual

MMS
®
Inspection SPG High 17
Getting started
4 Getting started
All the settings relevant to measuring
the height of surface profile of a sam-
ple
1
and the measured readings
themselves are saved in a file. Such a
file is called a batch.
In the batch, you define the measure-
ment procedure, e.g. whether the
specification limit is to be monitored
during the measurement or whether
the measured readings are to be
logged in measurement block groups.
The batch is always assigned the cal-
ibration 001 as a reference.
3 steps to the measurement
1. Switch on the gage,
2. Open a batch or create a new
one
Opening a batch, see
page 21
Creating a new batch, see
page 20
3. Take a measurement on the
sample, see Page 22
If the specified precision is not
achieved during the measure-
ment, you must recalibrate the
gage, see page 24.
1. In this manual, both a coated part of any
shape and size and a surface containing
multiple measurement locations are re-
ferred to as the sample.
After the gage is switched on,
the Measurement view (exam-
ple) for the batch that was
open at shutdown appears.
Batch 001
No 1 n 3/12
x 65.42 s 4.8
Min 60.2 Max 70.7
BLOCK
65.1
µm
75
50
After the gage is switched
on, the main menu appears
(example: Batch 001 is
open)
Batch 001
OK
Statistics
Batch Modify
Batches
Gage Settings