User Manual
Table Of Contents
MMS
®
Inspection SPG
13
Calibration For a correct measurement of the depth of surface profiles, the gage must record the two
extreme values "Zero" and "maximum depth" (= air value). This adjustment is carried out
by a calibration. If necessary, an adjustment to 1 to 2 further depth values is also possible.
General Features
Test method ASTM D4417, Method B, and magnetic induction method ISO 2178, ASTM D7091
Factory Calibration Each individual gage is factory calibrated at several reference points with the greatest care
to ensure the highest possible degree of trueness.
Data memory The memory content is preserved even when there is no voltage supply; subsequent viewing
of the measured single readings and evaluations
• Gage variant Start with memory capacity of max. 10,000 measured readings in 1 batch
• Gage variant High with memory capacity of 250,000 measured readings in 2500
batches
Evaluation Statistics
Display of mean value of all location values, standard deviation, min/ max values, range
and number of measured locations, number of measured readings lower/upper the set limit
values
Graphic Presentation
Run-Chart, showing the progress of stored readings
Probe Single tip axial probe with spring-loaded measuring tip built-in into gage
Measuring tip: 60° tip angle; Probe tip radius: 50 µm, hard metal
Probe tip replaceable by customer using the probe tip replacement kit 605-434
Quantity of measurements Before each measurement the probe tip has to be checked! After approximately 20,000
measurements the intact probe tip may show signs of wear and should be replaced.
Display of measurement acquisition Audible by a short beep and visual by colored illuminated LED; gage variant High: Addi-
tional by gage vibration
Display for limit monitoring • Limit violation: Audible by 2 short beeps and visual by red illuminated LED; gage variant
High: Additional by gage vibration
• Measured readings between the limits: Audible by 1 short beep and visual by green
illuminated LED; gage variant High: Additional by gage vibration
Languages German and English
Presettings for batches
Only available in gage variant High
Each new batch is created with a preset measurement unit and resolution for the displayed
measured value. You can adapt these presettings to your requirements. However, you can
also change the unit of measurement and the resolution for the measured value display at
any time in the batch that has already been created.
Display • Graphic display with automatic flipping measuring presentation view (deactivatable) to
read measurement results in many different gage positions
• Setting of brightness and contrast (definable for Office, Sunlight and Night)
Data transfer
Bluetooth/WiFi only available in gage variant
High
• USB: Data transfer of single readings to a PC, Data import to MSExcel via PC-Datex soft-
ware; You can gratis download the PC-Datex program from Fischer-Homepage
• Bluetooth/WiFi: Data transfer of measurements and data transfer of batches to App PHA-
SCOPE
®
PAINT; Creation and export of reports via App; You can gratis download the
App from Google Play Store and Apple App Store
USB port 2.0 Type C
• For service purpose
• For connection to PC for data transfer, max. cable length: 3 m (118 inches)