User guide
page 7 of 61
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011
Safety information!
1.1 General information
Read and comply with this manual before installing or operating the equipment.
Removeallpackingmaterial,removetheproductfromitspacking-box,removethe
protectivecoversfromtheinletandoutletportsandkeep,inspecttheequipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the product together with the order number and the
supplier’sinvoicenumber.Retainallpackingmaterialforinspection.
Do not use the equipment if it is damaged.
If the equipment is not used immediately, replace the protective covers. Store the
equipment in suitable conditions.
1.2 Intended use
☞ The pump and all system parts must not be used on humans or animals.
☞ Prevent any part of the human body from coming in contact with vacuum.
☞ Ensure that the individual components are only connected, combined and oper-
ated according to their design and as indicated in the instructions for use.
☞ Comply with notes on correct vacuum and electrical connections, see section
”Use and operation”.
• Thepumpsaredesignedforambient temperatures during operation between
+10°Cand+40°C.Checkthemaximumtemperaturesifinstallingthepumpina
cabinetorahousingandmakesureventilationisadequate.Installanexternal
automaticventilationsystemifnecessary.If pumpinghotprocessgasesmake
surethatthemaximumpermittedgasinlettemperature,whichdependsonsev-
eralparameterslikeinletpressureorambienttemperature(see”Technicaldata”),
isnotexceeded.
• Particles and dust must not be aspirated.
Use the equipment for the intended use only, i.e. for generation, control and meas-
urement of vacuum in vessels designed for that purpose.
1.3 Setting up and installing the equipment
➨ Equipment must be connected only to a suitably fused and protected electri-
cal supply and a suitable earth point. Failure to connect the motor to ground may
resultindeadlyelectricalshock.
ThemainscablemaybettedwithamouldedEuropeanIECplugoraplugsuit-
able for your local electrical supply. If the plug has been removed or has to be
removed, the cable will contain wires colour coded as follows: green or green and
yellow:earth;blueorwhite:neutral;brownorblack:live.
☞ Due to the high compression ratio of the pump, the pressure at the outlet port
might get higher than the maximum permissible pressure compatible with the
mechanical stability of the system.
☞ Do not permit any uncontrolled pressurizing(e.g.makesurethattheexhaust
pipelinecannotbecomeblocked).Ifthereisanexhaustisolationvalve,makesure
that you cannot operate the equipment with the valve closed. Risk of bursting!
NOTICE
NOTICE