➨ Connect hose to the connection at the inlet of the pump.
Tighten union nutrstby hand and thentightenone full
turn using open ended wrench.
➨ Assemble controller: Use Phillips screw driver to screw
bothscrewsthroughthexingplateinthesquarenutsin
the housing cover.
➨ Connect control cable to the rear side of the controller. Do
not cant when assembling connection!
➨ Tightenunionnut rst byhandand then tighten onefull
turn using open ended wrench.
If the pump does not achieve the ultimate vacuum:
- In case the diaphragms and valves have been replaced, a run-in period of several hours is required
before the pump achieves its ultimate vacuum.
- Incaseofunusualnoiseswitchoffpumpimmediatelyandcheckclampingdiscpositions.
Ifthespeciedultimatevacuumisnotachievedandifthisdoesnotchangeaftertherun-inperiod:
Checkhoseconnectors atpumphead for leaks.If necessary recheckvalveseats andpumpcham-
bers.