The controller itself is maintenance free.
Contaminationofthepressuretransducerordepositswillinuencetheaccuracyofmeasurement.
Attention: Neveruseaspikyorsharp-edgedtooltocleanthepressuretransducer.
Clean a contaminated pressure transducer as follows:
Disconnect vacuum connection.
Fillthemeasurementchamberwithasolvent(e.g.benzene)andallowsufcientcleaningtime.Ob-
serve all regulations concerning usage and disposal of solvents!
Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.
Rinse the gauge head chamber several times with alcohol in order to remove all solvent residues.
Allow the pressure transducer to dry.
Assemble vacuum connection.
Readjust the pressure transducer if necessary.
10 Cleaning the pressure transducer