The controller itself is maintenance free.
Contaminationofthepressuretransducerordepositswillinuencetheaccuracyofmeasurement.
Attention: Neveruseaspikyorsharp-edgedtooltocleanthepressuretransducer.
Clean a contaminated pressure transducer as follows:
➨ Disconnect vacuum connection.
➨ Fillthemeasurementchamberwithasolvent(e.g.benzene)andallowsufcientcleaningtime.Ob-
serve all regulations concerning usage and disposal of solvents!
➨ Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.
➨ Rinse the gauge head chamber several times with alcohol in order to remove all solvent residues.
➨ Allow the pressure transducer to dry.
➨ Assemble vacuum connection.
➨ Readjust the pressure transducer if necessary.