Attention: All error messages including a ashing warning triangle have to be conrmed (de-
leted) by pressing key START/STOP.
Fault Possible cause Remedy
Pump does not achieve
ultimate vacuum or nor-
mal pumping speed.
Wrongsettingsatcontroller?
Select function ”Pump down”
with speed set to ”HI” and
checkagain.
Centringringatasmallange
connection not correctly posi-
tionedorleakinthepipeline
orvacuumsystem?
Checkpumpdirectly-con-
nect controller directly at
pumpinlet-thencheckcon-
nection, pipeline and vacuum
system if necessary.
Long,narrowvacuumline?
Use lines with larger di-
ameter, length as short as
possible.
Pumphasbeenexposedto
condensate?
Allow pump to run for some
minutes with atmospheric
pressure at the inlet.
Deposits have been formed
insidethepump?
Clean and inspect the pump
heads.
Diaphragms or valves defec-
tive?
Replacediaphragmsand/or
valves.
Outgassing substances or
vapour generated in the proc-
ess?
Checkprocessparameters.
Pumptoohot?
Allow pump to cool down.
Determine and eliminate the
cause of overheating.
Pressure for automatic
switchingoffreached?
Adapt pressure for automatic
switching off, if necessary.