The controller is equipped with a serial interface (RS 232C, nine-pole Sub-D-plug).
☞ Plug-in or remove the cable (cable RS 232C) from the interface only if the equipment is switched off.
☞ The interface is not electrically isolated from the measuring circuit.
☞ Foroptimalelectromagneticcompatibilityassembleaninterfacelter(cat.no.:638235).
The controller is fully operable via the serial interface. Measuring results, preselections and the status of
the controller can be read at any time.
7.1 Setting of the interface
Set the interface parameters directly at the controller. The factory set values are underlined.
Editandconrmtheinterfaceparametersinthefunction”Conguration”inthemenu”RS-232”usingthe
selectionknob.
➨ Baud: 2400, 4800, 9600 or 19200
➨ Parity: 8-N-1, 7-O-1 or 7-E-1
➨ Handshake:Off,Xon-XofforRTS-CTS
➨ Remote: On or Off
➨ Timeout: Sending 1s, receiving 10s.
Inremotemode(RemoteOn,”PCsymbol”inthedisplay)allkeysatthecontrollerareinoperable.
To return to the manual operation of the controller set the controller to ”Remote off” in menu
conguration. Therefore switch off the controller. Switch on the controller and press selection
knob within 2s.
➨ Amaximumoftencommandspersecondispossible.
➨ Read commands and commands ”REMOTE”, ”CVC” and ”STORE” can be sent always. The sending
of other write commands is only possible, if ”Remote on” is selected.
➨ The commands have to be written in capital letters.
➨ Commandandparameterhavetobeseparatedbyablank.
➨ The string is terminated with <CR> or <CR><LF>.
➨ The answer of the controller is always terminated with <CR><LF>.
➨ Numerical values and parameters can be written without leading zeros.
➨ The answer of the controller is always with leading zeros.
7 Interface parameters