User guide
page 38 of 61
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011
The vacuum gauge was adjusted using factory standards, which are traceable
through regular calibration in an accredited laboratory (German Calibration service)
to the Germannationalpressurestandard.Dependingon the process and/or ac-
curacyrequirements,checktheadjustmentandreadjustifnecessary.Forreadjust-
ment, the device has to be adjusted both at atmospheric pressure as well as under
vacuumbutonly if thereference pressuresarecertainly known.The adjustment
modecanbe activated only iftheprocess control isinactive.Press key ”START/
STOP” if necessary. In the range between 20 and 700 mbar no adjustment is pos-
sible; ---- mbar is displayed.
Adjustment at atmospheric pressure
An adjustment at atmospheric pressure is only possible if the
pressure is higher than 700 mbar.
Ventilatethemeasurementconnectionofthecontrollerand/or
thevacuumsystem.Makesurethatthepressuretransducer
is at atmospheric pressure.
➨ In function ”Conguration” select program ”Adjustment” at
the controller.
➨ Usetheselectionknobtoadjustthereadingtothecurrent
atmospheric pressure.
➨ Presstheselectionknobtoconrm.
Adjustment under vacuum
An adjustment under vacuum is only possible if the pressure
is lower than 20 mbar absolute.
Evacuate the measurement connection of the controller to
a pressure < 0.1 mbar (e. g. by applying a good two-stage
rotary vane pump).
➨ In function ”Conguration” select program ”Adjustment” at
the controller.
☞ The reading is automatically adjusted to ”zero”.
➨ Presstheselectionknobtoconrm.
Adjustment at a reference pressure
Instead of adjustment under vacuum to a pressure < 0.1
mbar, adjustment to a precisely known reference pressure
within the range of 0 ..... 20 mbar is possible. Evacuate the
measurement connection of the controller to a pressure within
0 ..... 20 mbar
➨ In function ”Conguration” select program ”Adjustment” at
the controller.
☞ The reading is automatically adjusted to ”zero”.
➨ Usetheselectionknobtoadjustthedisplaytotherefer-
ence pressure at the vacuum line within the range of 0 ..... 20
mbar.
➨ Presstheselectionknobtoconrm.
1013 mbar
0 mbar
0...20 mbar
6 Readjustment of the controller
NOTICE
Note: To determine the actual atmospheric pressure, use an accurate barometer or get accurate read-
ing from the weather service, the next airport.......(take into account the difference in altitude between e.
g. airport and laboratory).
Note: Adjustment under vacuum with an actual pressure higher than 0.1 mbar reduces the accuracy of
the measurement. If the pressure is signicantly higher than 0.1 mbar, adjustment to a reference pres-
sure is recommended.
Note: The accuracy of the value of the reference pressure will directly affect the accuracy of the adjust-
ment. If the nominal ultimate vacuum of a diaphragm pump is used as reference vacuum, the accuracy
of the controller might be doubtful. The diaphragm pump may not achieve the specied value (due to
condensate, poor condition, failure of valves or diaphragm, leaks).